국산장비활용랩
|
Specification |
Hydraulic system |
Dual pistons in series pump with proprietary individual servo-controlled linear drive technology, |
Flow range |
0.001 - 2.000 mL/min(Max. 5ml/min) |
Flow precision |
≤ 0.075 % RSD or 0.005 min SD, whichever is greater |
Flow accuracy |
± 1% or ± 10 μL/min, whichever is greater |
Maximum operating pressure |
18,800 psi (1300 bar) |
Solvent selection |
2 solvents per pump module(A/B) |
Delay volume |
50 μL when using 35 μL mixer/Optional mixer : 100uL, 150uL |
Integrated degassing unit |
Integrated 2-channel degasser, with 480 μL chambers |
Communications |
LAN |
|
Specification |
Maximum data rate |
125 Hz (single wavelength detection) |
Noise |
-5 AU, at 254 nm (single wavelength detection) |
Linearity |
>2.0 AU upper limit |
Wavelength range |
190 – 900 nm |
Wavelength accuracy |
±1 nm, self-calibration with deuterium lines, verification with holmium oxide filter |
Wavelength precision |
<±0.1 nm |
Flow cells |
Standard: 2.4 µL volume, 10 mm cell path length and 1,000 psi Pressure maximum |
Communication |
LAN |
|
Specification |
Injection range |
Partial loopfill : 0 - 10 μL Full loop /μL pick-up: not applicable |
Precision Reproducibility (valid at 1.0 cP) |
RSD < 0.5% for partial loopfill injections, injection volumes > 5 μL at constant pressure |
Pressure range |
0 - 18,800 psi (0 – 1300 bar) |
Sample capacity |
96 x 2 mL vial (two trays default) |
Carry over |
0.05% with programmable needle wash |
Communications |
LAN |
|
Specification |
Temperature Range |
4℃~90℃ |
Temperature Stability |
0.05% |
No. of Columns |
Max. 3 ea of 15 cm columns |
Communications |
LAN |